E2105 – Peripheral Nerve Stimulator |
|
FOR THE NEUROMUSCULAR BLOCK MONITORING DURING GENERAL ANESTHESIA.
The E2105 allows the monitoring of neuromuscular block degree during general anesthesia procedures by the transcutaneous stimulation of a peripheral motor nerve and the observation of the corresponding muscle response. It generates the following stimulation modes used for the neuromuscular function evaluation:
-
TWT: Single Twitch

-
TET: Tetanic Stimulation
-
PTC: Post Tetanic Count
-
DBS: Double Burst Stimulation
-
TTOF: Train Of Four
These stimulation modes allow the assessment of:
-
The best time for intubation
-
The block degree during the anesthesia procedure
-
The need of an additional dose of the blocking agent
-
The presence of the residual block
Specifications:
|
Pulse amplitude: |
adjustable from 5 to 80 mA |
|
Pulse duration: |
200 µs |
|
Maximum pulse voltage: |
220 V |
|
Pulse shape: |
rectangular monophasic |
|
TWT mode: |
pulses at 1 pps |
|
TET mode: |
pulses at 50 pps |
|
DBS mode: |
two sequences of 3 pulses at 50 pps, the second 750 ms after the first. |
|
TOF mode: |
four pulses at 2 pps |
|
PTC mode: |
pulses at 50 pps during 5 s, no pulses during 3 s and 15 pulses at 1 pps/td> |
|
Power Source: |
one 9 V alkaline battery |
|
Dimensions: |
140 x 80 x 40 mm |
|
Weight: |
300 g (with battery) |
|